Model | HTSA-101D-W | HTSA-201D-W |
Power supply | 200V AC 3φ 3W 50/60Hz |
System | 2-zone and 3-zone with damper switching |
Performance *1 | Test area*2 | High temp. exposure temp. range | 1.Thermal shock cycle test +70°C to +150°C 2.Dew condensation cycle test -10°C to +100°C |
Low temp. exposure temp. range | -70°C to +10°C |
Temperature fluctuation | ± 1°C |
Humidity fluctuation | ± 5% rh |
High temperature and humidity chamber (stand-alone performance) | Upper limit pre heat. temp | +150°C |
Humidity range | 40 to 95% rh(at dew condensation cycle test) |
Temperature heat-up rate | Within 30 min. from -10°C to +100°C |
Temperature pull-down rate | Within 60 min. from +20°C to -10°C |
Low temperature chamber (stand-alone performance) | Lower limit pre-cool temperature | -75°C |
Temperature heat-up rate | Within 30 min. from -75°C to +10°C |
Temperature pull-down rate | Within 60 min. from ambient temp. to -75°C |
Temperature recovery performance *3 | Thermal shock cycle test | Recovery conditions | 3-zone | | Low temp. exposure: -65°C (30 min.) Ambient temp. exposure: Ambient temperature (10 min. ) High temp. exposure: +150°C (30 min.) | | Power supply : specified power supply Sensor position: Upstream of specimen | |
Specimen: Plastic mold IC 5kg | Specimen: Plastic mold IC 10kg |
Recovery time | High temp. recovery: within 10 min. (ambient temp. to +150°C), Low temp. recovery: within 10 min. (ambient temp. to -65°C) | High temp. recovery: within 16 min. (ambient temp. to +150°C), Low temp. recovery: within 16 min. (ambient temp. to -65°C) |
Dew condensation cycle test1 | Recovery conditions | 2-zone | | Low temp. exposure: +5°C (20 min.) High temp. and humidity exposure: +25°C 90% (20 min.) | | Power supply : specified power supply Sensor position: Upstream of specimen | |
Specimen: PCB 2kg | Specimen: PCB 5kg |
Recovery time*4 | High temp. recovery: within 20 sec, (+5°C→ +25°C 90%) Low temp. recovery: within 3 min. (+25°C 90% rh → +5°C) | High temp. recovery: within 20 sec, (+5°C→ +25°C 90%) Low temp. recovery: within 5 min. (+25°C 90% rh → +5°C) |
Dew condensation cycle test2 | Recovery conditions*5 | 2-zone | | Low temp. exposure: -30°C(60 min.) High temp. and humidity exposure: +25°C 95%(60 min.) *Preheat temp.25°C 95% | | Power supply : specified power supply Sensor position: Upstream of specimen Specimen: none | |
Recovery time*4 | High temp. and humidity recovery: within 5 min.(-30°C→ +25°C 95%) Low temp. and humidity recovery: within 5 min. (+25°C 95% rh → -30°C) |
Specimen basket withstand load | 5kg(Evenly distributed load) |
Test area dimensions | W650 × H460 × D370mm | W650 × H460 × D670mm |
Chamber outer dimensions | W1670 × H1900 × D1570mm (excluding protrusions) | W1670 × H1900 × D1870mm (excluding protrusions) |
Chamber weight | About 1300 kg | About 1550 kg |